- Regulatory Status
- RUO
- Other Names
- Retrieve-All Antigen Unmasking Solution
- Previously
-
Signet Catalog# 1910
Covance Catalog# SIG-31910
- Ave. Rating
- Submit a Review
- Product Citations
- publications
Cat # | Size | Price | Quantity Check Availability | Save | ||
---|---|---|---|---|---|---|
927901 | 50 mL | 272€ |
Product Details
- Formulation
- Antigen Unmasking Solution.
- Storage & Handling
- Store at room temperature.
- Application
-
IHC
- Recommended Usage
-
The optimal antigen unmasking condition is a function of temperature (90 to 100°C) and incubation time (5-30 minutes) and should be established by each investigator.
The Retreive-All working solution should be prepared by mixing 10 mL of 100x Retrieve-All into 990mL of water.
Protocol
1. Deparaffinize sections, rehydrate through graded alcohols, DI water, into PBS for 5 min.
2. Preheat Retrieve-All buffer to 92°C*.
3. Place slides into the preheated Retrieve-All buffer.
4. Incubate for 10 minutes at 92°C.
5. Remove the container with slides and let it cool at room temperature for 5-10 minutes.
6. Rinse in distilled water or PBS.
7. Proceed with usual immunohistochemical staining. - Application Notes
-
*Concentrated buffers (100x) may show minimal precipitation after long term storage without any effect to the HIER.
Retrieve-All is an antigen un-masking solution available in the following three pH formulas for use in heat induced un-masking procedures:
Retrieve-All 1 (Universal pH 8)
Retrieve-All 2 (Basic pH 10)
Retrieve-All 3 (Acidic pH 4.8)
Antigen Details